{"schema":"/api/v1/schema/cve.json","api_version":"1","id":"CVE-2017-5149","published":"2017-02-13T22:59:00.303000Z","last_modified":"2026-06-17T01:20:03.097000Z","status":"Modified","cvss":{"version":"3.1","score":8.9,"severity":"HIGH","vector":"CVSS:3.1/AV:N/AC:H/PR:N/UI:N/S:C/C:L/I:H/A:H","source":"nvd@nist.gov","primary":true},"auth":false,"cwes":["CWE-476"],"cpes":[{"criteria":"cpe:2.3:o:abbott:merlin\\@home_firmware:*:*:*:*:*:*:*:*","part":"o","vendor":"abbott","product":"merlin@home_firmware","version_end_including":"8.0"}],"cpes_truncated":false,"vendors":["abbott"],"products":["abbott/merlin-home_firmware"],"cat":"firmware","platform":"appliance","eol":false,"packages":[],"severity":"high","priority_score":68,"links":{"html_es":"/es/cve/CVE-2017-5149/","html_en":"/en/cve/CVE-2017-5149/","nvd":"https://nvd.nist.gov/vuln/detail/CVE-2017-5149","cve_org":"https://www.cve.org/CVERecord?id=CVE-2017-5149"},"untrusted":{"vendor_name":"Abbott","product_name":"Merlin@home Firmware","desc_en":"An issue was discovered in St. Jude Medical Merlin@home, versions prior to Version 8.2.2 (RF models: EX1150; Inductive models: EX1100; and Inductive models: EX1100 with MerlinOnDemand capability). The identities of the endpoints for the communication channel between the transmitter and St. Jude Medical's web site, Merlin.net, are not verified. This may allow a man-in-the-middle attacker to access or influence communications between the identified endpoints.","affected":[],"references":[{"url":"http://www.securityfocus.com/bid/95331","tags":["Third Party Advisory","VDB Entry"]},{"url":"https://ics-cert.us-cert.gov/advisories/ICSMA-17-009-01A","tags":["Mitigation","Third Party Advisory","US Government Resource"]}]}}