Live threats

All CVEs / Bd

CVE-2016-9355Bd Alaris 8015 Pc Unit

Medium 5.3

Priority
32
CVSS 3.0
5.3
EPSS
0.5%
probability of exploitation in 30 days · percentile 37
Auth required
Yes
Published
Feb 13, 2017
Updated
Jun 17, 2026
Category
Application · Cross-platform
Weakness (CWE)
CWE-255
Vector
CVSS:3.0/AV:P/AC:L/PR:N/UI:N/S:C/C:H/I:N/A:N

What happens

An issue was discovered in Becton, Dickinson and Company (BD) Alaris 8015 Point of Care (PC) unit, Version 9.5 and prior versions, and Version 9.7. An unauthorized user with physical access to an Alaris 8015 PC unit may be able to obtain unencrypted wireless network authentication credentials and other sensitive technical data by disassembling an Alaris 8015 PC unit and accessing the device's flash memory. Older software versions of the Alaris 8015 PC unit, Version 9.5 and prior versions, store wireless network authentication credentials and other sensitive technical data on the affected device's removable flash memory. Being able to remove the flash memory from the affected device reduces the risk of detection, allowing an attacker to extract stored data at the attacker's convenience.

Original NVD text.

What to do

  1. Apply the vendor's security update for Alaris 8015 Pc Unit.

Affected products (CPE)

VendorProductVersions
bdalaris_8015_pc_unit≤ 9.5
bdalaris_8015_pc_unit9.7

Official sources

References

Links published by the sources. Open them with care.