CVE-2017-12721Smiths-medical Medfusion 4000 Wireless Syringe Infusion Pump
Medium 5.9 No auth needed
- Priority
- 50
- CVSS 3.0
- 5.9
- EPSS
- —
- Auth required
- No
- Published
- Feb 15, 2018
- Updated
- Jun 17, 2026
- Category
- Operating system · Cross-platform
- Weakness (CWE)
- CWE-295
- Vector
CVSS:3.0/AV:N/AC:H/PR:N/UI:N/S:U/C:H/I:N/A:N
What happens
An Improper Certificate Validation issue was discovered in Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump, Version 1.1, 1.5, and 1.6. The pump does not validate host certificates, leaving the pump vulnerable to a man-in-the-middle (MITM) attack.
What to do
- Apply the vendor's security update for Medfusion 4000 Wireless Syringe Infusion Pump.
- It can be exploited without authentication: if you cannot patch now, restrict Internet access to the affected service.
Affected products (CPE)
| Vendor | Product | Versions |
|---|---|---|
| smiths-medical | medfusion_4000_wireless_syringe_infusion_pump | 1.1 |
| smiths-medical | medfusion_4000_wireless_syringe_infusion_pump | 1.5 |
| smiths-medical | medfusion_4000_wireless_syringe_infusion_pump | 1.6 |