CVE-2019-25721Dräger Infinity M300
High 7.1 No auth needed
Follow: this CVE · Dräger · Dräger Infinity M300 We alert you by email or Slack (account required).
- Priority
- 58
- CVSS 4.0
- 7.1
- EPSS
- —
- Auth required
- No
- Published
- Jun 2, 2026
- Updated
- Jul 22, 2026
- Category
- Application · Cross-platform
- Weakness (CWE)
- CWE-400
- SSVC (CISA)
- Exploitation: none · Automatable: no · Technical impact: partial
- Vector
CVSS:4.0/AV:A/AC:L/AT:N/PR:N/UI:N/VC:N/VI:N/VA:H/SC:N/SI:N/SA:N/E:X/CR:X/IR:X/AR:X/MAV:X/MAC:X/MAT:X/MPR:X/MUI:X/MVC:X/MVI:X/MVA:X/MSC:X/MSI:X/MSA:X/S:X/AU:X/R:X/V:X/RE:X/U:X
What happens
Dräger Infinity M300 patient worn monitors with software version VG2.3.1 and earlier contain a network-based denial of service vulnerability that allows network-adjacent attackers to repeatedly trigger device reboots by sending malicious requests over the Infinity Network. Attackers can exploit this vulnerability to force the device into a fail state requiring manual restart, causing loss of wireless connectivity and interruption of patient monitoring functionality.
What to do
- Apply the vendor's security update for Infinity M300.
- It can be exploited without authentication: if you cannot patch now, restrict Internet access to the affected service.
Affected products
- Dräger · Infinity M300
As reported by the CNA that published the CVE. NVD has not added CPEs with versions yet.
Official sources
References
Links published by the sources. Open them with care.