CVE-2018-10596Medtronic 2090 Carelink Programmer Firmware
Alta 7.1
Seguir: esta CVE · Medtronic · Medtronic 2090 Carelink Programmer Firmware Te avisamos por email o Slack (requiere cuenta).
- Prioridad
- 43
- CVSS 3.1
- 7.1
- EPSS
- —
- Requiere autenticación
- Sí
- Publicada
- 3 jul 2018
- Actualizada
- 17 jun 2026
- Categoría
- Firmware · Equipo de red
- Vector
CVSS:3.1/AV:N/AC:H/PR:L/UI:R/S:U/C:H/I:H/A:H
Qué pasa
Medtronic 2090 CareLink Programmer
uses a virtual private network connection to securely download updates. It does not verify it is still connected to this virtual private network before downloading updates. The affected products initially establish an encapsulated IP-based VPN connection to a Medtronic-hosted update network. Once the VPN is established, it makes a request to a HTTP (non-TLS) server across the VPN for updates, which responds and provides any available updates. The programmer-side (client) service responsible for this HTTP request does not check to ensure it is still connected to the VPN before making the HTTP request. Thus, an attacker could cause the VPN connection to terminate (through various methods and attack points) and intercept the HTTP request, responding with malicious updates via a man-in-the-middle attack. The affected products do not verify the origin or integrity of these updates, as it insufficiently relied on the security of the VPN. An attacker with remote network access to the programmer could influence these communications.
Qué hacer
- Aplica la actualización de seguridad del fabricante para 2090 Carelink Programmer Firmware.
Productos afectados (CPE)
| Fabricante | Producto | Versiones |
|---|---|---|
| medtronic | 2090_carelink_programmer_firmware | todas |
Fuentes oficiales
Referencias
Enlaces publicados por las fuentes. Ábrelos con criterio.
- global.medtronic.com
- www.cisa.gov
- ics-cert.us-cert.gov Third Party Advisory US Government Resource