Amenazas en vivo

Todas las CVE / Medtronic

CVE-2018-10596Medtronic 2090 Carelink Programmer Firmware

Alta 7.1

Prioridad
43
CVSS 3.1
7.1
EPSS
—
Requiere autenticación
Sí
Publicada
3 jul 2018
Actualizada
17 jun 2026
Categoría
Firmware · Equipo de red
Debilidad (CWE)
CWE-923, CWE-200
Vector
CVSS:3.1/AV:N/AC:H/PR:L/UI:R/S:U/C:H/I:H/A:H

Qué pasa

Medtronic 2090 CareLink Programmer uses a virtual private network connection to securely download updates. It does not verify it is still connected to this virtual private network before downloading updates. The affected products initially establish an encapsulated IP-based VPN connection to a Medtronic-hosted update network. Once the VPN is established, it makes a request to a HTTP (non-TLS) server across the VPN for updates, which responds and provides any available updates. The programmer-side (client) service responsible for this HTTP request does not check to ensure it is still connected to the VPN before making the HTTP request. Thus, an attacker could cause the VPN connection to terminate (through various methods and attack points) and intercept the HTTP request, responding with malicious updates via a man-in-the-middle attack. The affected products do not verify the origin or integrity of these updates, as it insufficiently relied on the security of the VPN. An attacker with remote network access to the programmer could influence these communications.

Texto original de NVD, en inglés. La traducción y el resumen en español llegan pronto.

Qué hacer

  1. Aplica la actualización de seguridad del fabricante para 2090 Carelink Programmer Firmware.

Productos afectados (CPE)

FabricanteProductoVersiones
medtronic2090_carelink_programmer_firmwaretodas

Fuentes oficiales

Referencias

Enlaces publicados por las fuentes. Ábrelos con criterio.