CVE-2015-1012Pfizer Lifecare Pca Infusion System Firmware
High 7.5 No auth needed
Follow: this CVE · Pfizer · Pfizer Lifecare Pca Infusion System Firmware We alert you by email or Slack (account required).
- Priority
- 60
- CVSS 3.0
- 7.5
- EPSS
- 0.8% probability of exploitation in 30 days · percentile 52
- Auth required
- No
- Published
- Mar 25, 2019
- Updated
- Jun 17, 2026
- Category
- Firmware · Network appliance
- Vector
CVSS:3.0/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:N/A:N
What happens
Wireless keys are stored in plain text on version 5 of the Hospira LifeCare PCA Infusion System. According to Hospira, version 3 of the LifeCare PCA Infusion System is not indicated for wireless use, is not shipped with wireless capabilities, and should not be modified to be used in a wireless capacity in a clinical setting. Hospira has developed a new version of the PCS Infusion System, version 7.0 that addresses the identified vulnerabilities. Version 7.0 has Port 20/FTP and Port 23/TELNET closed by default to prevent unauthorized access.
What to do
- Apply the vendor's security update for Lifecare Pca Infusion System Firmware.
- It can be exploited without authentication: if you cannot patch now, restrict Internet access to the affected service.
Affected products (CPE)
| Vendor | Product | Versions |
|---|---|---|
| pfizer | lifecare_pca_infusion_system_firmware | ≤ 5.0 |
Official sources
References
Links published by the sources. Open them with care.
- ics-cert.us-cert.gov Mitigation Third Party Advisory US Government Resource