CVE-2015-1012Pfizer Lifecare Pca Infusion System Firmware
Alta 7.5 Sin autenticación
- Prioridad
- 60
- CVSS 3.0
- 7.5
- EPSS
- —
- Requiere autenticación
- No
- Publicada
- 25 mar 2019
- Actualizada
- 17 jun 2026
- Categoría
- Firmware · Equipo de red
- Vector
CVSS:3.0/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:N/A:N
Qué pasa
Wireless keys are stored in plain text on version 5 of the Hospira LifeCare PCA Infusion System. According to Hospira, version 3 of the LifeCare PCA Infusion System is not indicated for wireless use, is not shipped with wireless capabilities, and should not be modified to be used in a wireless capacity in a clinical setting. Hospira has developed a new version of the PCS Infusion System, version 7.0 that addresses the identified vulnerabilities. Version 7.0 has Port 20/FTP and Port 23/TELNET closed by default to prevent unauthorized access.
Qué hacer
- Aplica la actualización de seguridad del fabricante para Lifecare Pca Infusion System Firmware.
- Se explota sin autenticación: si no puedes parchear ya, limita el acceso desde Internet al servicio afectado.
Productos afectados (CPE)
| Fabricante | Producto | Versiones |
|---|---|---|
| pfizer | lifecare_pca_infusion_system_firmware | ≤ 5.0 |
Fuentes oficiales
Referencias
Enlaces publicados por las fuentes. Ábrelos con criterio.
- ics-cert.us-cert.gov Mitigation Third Party Advisory US Government Resource